Description
The BioPlex 2200 System is a fully automated, random access multiplex system that allows clinical laboratories to maximize efficiency for autoimmune, infectious disease, and specialty testing.
Specifications
Physical dimensions
Height: 1.3 m
Width: 1.3 m
Depth: 0.9 m
Weight: 468 kg
Performance
Maximum Throughput*: Up to 1,400 patient results per hour
Sample Capacity: Up to 100 samples per hour
Time to First Result: As little as 45 minutes (panel dependent)
Sample Tubes: Any CLSI primary sample tubes
Sample Tray: 10 racks, 5 samples each
Sample Volume: As little as 5 µL
STAT Capability: Continuous access
Walkaway Capacity: 8 hours of continuous use
Operational Mode: Random and continuous access, priority (STAT) loading
Operator Interfaces: Touchscreen monitor, keyboard, trackball mouse
Clot Detection: Yes
Quality Control: Levey-Jennings, Westgard Rules
Sample Barcode Acceptable Symbologies: Code 128, Codabar, Code 39, Interleaved 2 of 15
Sample Types: Serum, plasma, or CSF (for more information, refer to BioPlex 2200 Assay Inserts)
Sample Dilutions: 1:4, 1:8, 1:10, 1:16, 1:32, 1:64, and 1:100 (panel dependent)
Access During Use: Continuous sample input/output, consumables replenishment, and waste disposal without interruption of sample processing
LIS Connectivity: Bidirectional: ASTM Standard
Total Laboratory Automation (TLA): Complies with CLSI Auto5A interface standard
* ANA Throughput
Power Requirements
Input Voltage: 100-240 VAC at 50-60 Hz, 1000 W
UPS: Provided
Water Requirements
Water Quality: CLSI Type II or better
Direct Plumbing: Maximum pressure of 50 kPa (7.25 psi)
Installation Requirements
Crate Requirements
Rear: 1.5 m
Left: 1.2 m
Right: 1.2 m
Front: 3.0 m
Vertical: 1.8 m
Instrument Operational Clearance
Rear: 0.2 m
Left: 1.0 m
Right: 0.2 m
Front: 1.0 m
Vertical: 1.8 m



Reviews