Description
The ST-B60/80PV series is a type B sterilizer that uses steam as a medium. It achieves sterilization of items by repeatedly extracting vacuum and filling the sterilization chamber with steam to create a certain level of vacuum before introducing saturated steam, reaching the set pressure and temperature to achieve the purpose of sterilization. It is suitable for use in medical and health care departments such as stomatology, ophthalmology, operating rooms, supply rooms, and analysis rooms for sterilizing solid instruments, type A hollow instruments, implantable instruments, dressings and fabrics, rubber catheters, and liquid items.
The non-liquid program is suitable for the sterilization of surgical instruments, solid bare instruments, packaged instruments, rubber loads, etc. The liquid program is suitable for the sterilization of liquids such as water and culture media.
Benchtop Pulse Vacuum Autoclave Specifications:
Model ST-B60PV ST-B80PV
Capacity 60L 80L
Chamber Size(mm) φ391×521 φ391×671
Designed Pressure -0.1/0.3Mpa
Sterilization Temperature 150℃
Wokring Temperature 121℃/134℃
Water Tank Capacity(L) 14
Chamber Material SUS304
Noise ≤65dB
Consumption 5.8kW 6.2kW
Power Supply 220V/110V, 50/60HZ
External Size(WxDxH)(mm) 990x804x675 1140x805x675
Package Size(WxDxH)(mm) 1130x960x940 1330x960x940
N.W./G.W. (kg) 175/215 185/230
Benchtop Pulse Vacuum Autoclave Features:
Control System
Display:The LCD touch screen can display temperature, pressure, time, operating status, fault alarm, equipment operation curve and other information in real time, making the sterilization information more intuitive and clear, and convenient for users to observe the operating status of the equipment
Process control
The whole process of preheating, replacement, heating, sterilization, exhaust and vacuum drying is automatically operated; it uses steam as the medium, adopts an efficient vacuum pumping system, and repeatedly pumps vacuum to achieve the effect of sterilization and drying.
Multi-point Temperature Control Technology
Non-sealed liquid programs use mobile temperature sensors to measure tempera-ture, which can detect the internal temperature of the liquid in real time, making the internal temperature of the liquid more intuitive and clear.
Insulation Function
The insulation function can be set as needed to achieve liquid culture medium sterilization, culture medium dissolution-insulation function.
Safety Function
With over-temperature automatic protection device, anti-dry burning protection device, over-pressure automatic release device, over-current protection device, electronic circuit safety device
Self-calibration Function
It has a complete background self-calibration system to realize the calibration of system parameters such as pressure and temperature. Without disassembling the instrument, the authorized tool can be used for on-site adjustment.
Independently Installed Pressure Sensors
An independent pressure transmitter is installed in each of the inner chamber and evaporator to ensure that the equipment pressure is within the safe control range
Water Injection and Drainage Method
Automatic water injection, automatic drainage
Pressure Gauge: Range
-0.1~0.5MPa, Accuracy level: 1.6
Filter
Manually removable filter, easy to clean, more user-friendly
Sterilization Program:
Encapsulation Program, Rubber Program, Dressing Program, Liquid Program, Liquid customization, Solid customization, as well as B&D testing, Vacuum testing Program, Medium Sterilization & Dissolution







HORIBA Pentra C200 Hematology Analyzer
Hycel HY-LYTE 200 Electrolyte Analyzer with ISE
HORIBA Yumizen H500 OT Hematology Analyzer
Horiba Yumizen G200 Semi Automated Coagulation Analyzer
Hycel HYCOUNT 5 XL Hematology Analyzer
HORIBA ABX ABX Pentra DF 120 SPS Hematology Analyzer
HORIBA ABX Pentra 120 Retic SPS Hematology Analyzer
Reviews
There are no reviews yet.