Description
Key Feature
Feature Icon
Pressure
0-50/69MPa
Feature Icon
Temperature
-196℃ to 1100℃
Feature Icon
Main Function
High pressure Isotherm, PCT, Multi Component adsorption analysis, Automatic Cyclic Test
Feature Icon
Optional Function
Kinetics, TPD, TPA,
PCT with larger sample volume (100g level)
Feature Icon
Full Automation
Automatically switches from pre-treatment to testing
Feature Icon
Thermostatic design
The analyzer pipeline temperature is 40.0℃ ±0.1℃.
Feature Icon
Sample cell options
Stainless steel sample cell :2ml/ 5ml/25ml;
Core sample
cell:Φ25.4×50mm/Φ50×110mm/Φ38.1×100mm
Feature Icon
High-pressure design
All stainless steel gas circuit system, the joints are hard-connected with VCR, which improves the stability of the instrument.
Feature Icon
Transducers
High precision with 0.01%FS, long term 0.025%FS
Specifications
Specifications Details
Main Function High pressure adsorption and desorption isotherm
Multi Component high pressure adsorption analysis High pressure PCT
Automatic Cyclic Test
Optional Function Isobaric adsorption/desorption kinetics
Temperature programmed desorption (TPD)
Temperature programmed adsorption (TPA)
Vacuum desorption
PCT test with larger sample volume (100g level)
Analysis station 1/2
Degassing station 1/2
Pressure 0-20MPa
Temperature -196 to 1100℃
Transducer 2x pcs 10bar and 200bar with 0.01% full scale
Vacuum 10-2 Pa by oil pump
10-8 Pa by molecular pump
Analysis gas H2、CH4、Ar、CO2、N2、Kr、O2、C2H6、C2H4、He etc.
Repeatability ±2% by standard sample
Sample cell Stainless steel sample cell :2ml/ 5ml/25ml;
Core sample cell:Φ25.4×50mm/Φ50×110mm/Φ38.1×100mm
Self-sealing sample tube
Automation Automatically switches from pre-treatment to testing after setup
Instrument thermostatic temperature The inside of the instrument is in the same air bath environment, the thermostatic temperature is 40.0℃, and the temperature control accuracy is ±0.1℃.
High-pressure gas pipeline All stainless steel gas circuit system, the joints are hard-connected with VCR, which improves the stability of the instrument.
Filling rod with filter Standard sample tube filling rod with filtering function can effectively reduce the free space volume of the sample tube and prevent the sample from flying and contaminating the instrument pipelines.